The gamma camera

Radionuclide imaging Free lesson Lesson 30 of 90

NaI(Tl) crystal, PMT array, position and energy signals, planar operation, positioning and contouring, detector care (temperature, mechanical)

0:00 / 0:00

Do you like iRad Shots?

Tap a star

INDEX

The shot

In one line

A gamma camera takes each gamma photon through a collimator into a thin crystal, where it becomes a flash of light. Photomultiplier tubes then find where the flash was and its energy. Counted one by one, the photons build the image.

Know this
  • The photon first meets the collimator, a lead plate full of holes. Only photons travelling along the holes get through, so each shows where it came from.
  • Behind it lies one large, thin crystal of sodium iodide activated with thallium. Absorbing the photon, it gives a flash of a few thousand blue light photons.
  • A glass light guide spreads the light to an array of photomultiplier tubes, each turning its share into an electrical pulse.
  • The tubes nearest the flash get the most light. Their shares give the position signals, X and Y. Their sum, Z, is the energy signal, proportional to the photon's energy.
  • A thin crystal stops most 140 keV photons. A thicker one stops more high-energy photons, but the light spreads further, blurring the position.
  • Each accepted photon adds one count to its pixel. A planar image is a map of counts: the more counts, the less noise.
  • Place the detector as close to the patient as possible and follow the body's contour. Resolution worsens with distance.
  • The crystal is fragile. It absorbs moisture, so it is sealed, and it cracks with knocks or rapid changes of temperature.
Numbers
  • Crystal: about 9.5 mm thick, about 40 × 50 cm.
  • 9.5 mm stops about 90% of 140 keV photons.
  • About 40 light photons per keV: about 5,000 at 140 keV.
  • About 40 to 90 photomultiplier tubes.
  • At 140 keV: intrinsic resolution about 3.5 mm, energy resolution about 10%.
Exam traps
  • Position comes from each tube's share of the light, so it is far finer than the tube spacing.
  • The energy signal is the sum of all the tubes, not the brightest tube.
  • A thicker crystal improves sensitivity at high energies but worsens intrinsic resolution.
  • Thallium is only the activator. Iodine stops the gamma rays.

Sketch

Hand-drawn sketch: The gamma camera

Check yourself

As in the exam: mark each item true or false. Any number can be true.

1. Concerning the crystal of a gamma camera:

  1. aThe number of light photons from one gamma ray is proportional to the energy deposited.
  2. bIt is a mosaic of small separate crystals, one for each photomultiplier tube.
  3. cIt is typically about 9.5 mm thick.
  4. dIt emits light in the red part of the spectrum.
  5. eIt is sodium iodide activated with thallium.

2. Regarding the thickness of the crystal:

  1. aA thin crystal has poor sensitivity for 364 keV photons.
  2. bA thicker crystal improves intrinsic spatial resolution.
  3. cCrystal thickness has no effect on sensitivity.
  4. dSome cameras use a thicker crystal for high-energy imaging.
  5. eAbout 90% of 140 keV photons are stopped in 9.5 mm of sodium iodide.

3. Concerning the photomultiplier tubes and signals of a gamma camera:

  1. aThe position signals come from each tube's share of the light.
  2. bThe energy signal is the sum of all the tubes' signals.
  3. cA head typically holds about 40 to 90 photomultiplier tubes.
  4. dThe intrinsic resolution is limited to the diameter of one tube.
  5. eOnly the tube directly in front of the flash produces a signal.

4. Regarding planar imaging with a gamma camera:

  1. aThe detector should be kept well away from the patient to improve resolution.
  2. bA planar image shows the depth of the activity.
  3. cEach accepted photon adds one count to a pixel.
  4. dImage noise falls as the number of counts rises.
  5. eResolution with a parallel-hole collimator worsens with distance from the face.

5. Concerning care of the gamma camera detector:

  1. aThe gamma rays enter the crystal through its glass window.
  2. bSodium iodide is hygroscopic.
  3. cThe crystal is unaffected by rapid changes of temperature.
  4. dThe collimator or a cover is kept on the face of the detector.
  5. eA cracked crystal can be detected only by opening the detector.

More questions in the question bank
The full explanation about 5 min to read

What the camera has to do

A radiopharmaceutical spreads through the patient, and gamma photons leave the body in every direction. The gamma camera must answer two questions about each photon it catches: where did it come from, and was it a photon that left the patient without being scattered? It answers them one photon at a time, and the image is the record of the answers. This lesson follows one photon of technetium-99m, 140 keV, through the camera head.

The collimator

A photon could strike the crystal from any angle, so on its own the crystal would record a blur. The collimator, a thick lead plate pierced by thousands of parallel holes, lets through only photons travelling along the holes, at right angles to the face. A photon that reaches the crystal at a given point therefore came from the part of the patient directly in front of that point. The price is that almost every photon is stopped by the lead, which is why the camera is so much less sensitive than an X-ray detector. The kinds of collimator, and the balance between resolution and sensitivity, are the next lesson.

The crystal

Behind the collimator lies a single large crystal of sodium iodide activated with thallium, NaI(Tl), about 9.5 mm thick, three-eighths of an inch, and on a modern rectangular head about 40 by 50 cm. Iodine, atomic number 53, absorbs gamma rays well, mostly by the photoelectric effect at 140 keV; thallium is only a trace, added so that the crystal gives out light efficiently (band theory and activators are in the scintillator lesson). The absorbed energy reappears as a flash lasting a fraction of a microsecond: about 40 photons of blue light, 415 nanometres, for each keV, so about 5,000 for a 140 keV photon. The number of light photons is proportional to the energy absorbed, and that is what lets the camera measure energy.

It is sealed in a thin aluminium case, which the gamma rays cross easily, with a glass window at the back to let the light out.

Why thin?

About 90% of 140 keV photons are stopped in 9.5 mm of sodium iodide, so a thicker crystal would add little at this energy. It would also blur the image: the flash spreads out as it travels to the tubes, and the further it travels the wider the pool of light and the less precisely its centre can be found. For higher-energy photons, such as iodine-131 at 364 keV, a thin crystal lets many pass straight through, and sensitivity falls. Some cameras use a thicker crystal, for example 15.9 mm, five-eighths of an inch, accepting slightly poorer resolution for better sensitivity at high energies.

Light guide and photomultiplier tubes

The light passes through a glass light guide, optically coupled to the crystal, into an array of photomultiplier tubes packed across the back: typically about 40 to 90 in a head, round or hexagonal, each a few centimetres across. In each tube, light strikes a photocathode and releases electrons, which are multiplied at a chain of dynodes, roughly a million times, into a pulse large enough to measure. The light guide spreads the light so that several tubes always share a flash, even one directly in front of a tube.

Where, and how much

A flash shares its light among the tubes according to distance: the tubes nearest it receive the most, the more distant ones less. The positioning electronics, analogue in older cameras and digital in modern ones, weight each tube's signal by the tube's position and add them up, giving the X and Y position signals. Each is divided by the total, so that the position does not depend on the brightness of the flash. Because the position is worked out from the share of the light, it is found to within a few millimetres, far finer than the spacing of the tubes: the intrinsic resolution is about 3.5 mm at 140 keV.

The sum of all the tubes' signals is the energy signal, Z, proportional to the energy deposited. A photon scattered in the patient has lost energy and arrives with a smaller Z. Pulse-height analysis accepts only photons whose Z lies in a window around the photopeak; the energy resolution, about 10% at 140 keV, decides how well scatter can be rejected. Windows, uniformity and quality control are covered later.

The planar image

Each accepted photon adds one count to the pixel of a matrix at its X and Y. Over a few minutes, counts build up where the activity is concentrated, and the planar image is that map of counts: a two-dimensional projection of the activity, with depth lost. Its noise falls as counts rise, which is why acquisitions are set by counts or time (acquisition modes, and SPECT for depth, come later).

Positioning the patient

With a parallel-hole collimator, resolution worsens steadily with distance from the collimator face, while sensitivity hardly changes. So the detector goes as close to the patient as possible, and for whole-body and tomographic imaging it follows the body's contour rather than a wide circle. A gap between patient and camera costs sharpness rather than counts.

Care of the detector

The crystal is brittle and is easily cracked by a knock, so the collimator, or a protective cover, stays on the face, and collimator changes are done with care. Sodium iodide is hygroscopic: it absorbs moisture from the air and yellows, which is why it is sealed; a broken seal ruins it. It also cracks with rapid changes of temperature, so the room is kept at a steady temperature and the head is never warmed or cooled quickly, for example by sunlight or a heater blowing on it. Damage shows as defects on the uniformity flood.

Exam traps

One: position comes from each tube's share of the light, so it is far finer than the spacing of the tubes. Two: the energy signal is the sum of all the tubes, proportional to the photon's energy. Three: a thicker crystal improves sensitivity at high energies but worsens intrinsic resolution. Four: thallium is the activator; iodine stops the gamma rays.