CTDI100 / CTDIw / CTDIvol, DLP, effective dose from conversion factors, why CTDIvol is not the patient's dose (size-specific estimates), typical values
A CT dose report gives CTDIvol, the scanner's output in a standard acrylic cylinder, and DLP, that times the length scanned. Neither is the patient's dose.
As in the exam: mark each item true or false. Any number can be true.
1. Concerning the CT dose index:
2. Regarding CTDIvol:
3. Concerning the dose-length product and effective dose:
4. Regarding CTDIvol and patient size:
5. Concerning UK national diagnostic reference levels for adult CT:
Every CT scanner shows two dose numbers before it scans and records them afterwards: the volume CT dose index, CTDIvol, in milligray, and the dose-length product, DLP, in milligray centimetres. Both describe the scanner's output; neither is measured on the patient. Gray, sievert and effective dose are lesson L40.
The measurement is made in a phantom: a cylinder of acrylic (polymethyl methacrylate) about 14 to 15 cm long, 16 cm across for head protocols and 32 cm for body protocols. Each has five holes along its length: one at the centre and four round the edge, just under the surface.
Into a hole goes a pencil ionisation chamber 100 mm long. The scanner makes one axial rotation with the couch still. The beam is only a few centimetres wide, but scatter spreads dose along the patient's axis, z, in tails either side. The chamber collects that rotation's dose along its whole length, tails included, and dividing by the beam width gives CTDI100. Tails beyond the chamber are missed.
The dose is not even across the slice. In the body phantom the edge typically reads about twice the centre, which is shielded by the acrylic around it; in the head phantom they are closer. The weighted index is CTDIw = one third of the centre reading + two thirds of the edge reading (the average of the four edge holes). The thirds roughly match the areas the centre and the edge stand for, so CTDIw stands for the average across the slice.
If the couch moves further than the beam width each turn, the dose is spread more thinly, so CTDIvol = CTDIw ÷ pitch (pitch is lesson L54). It is the average dose within the scanned volume of that phantom for that technique, and it is the number the scanner displays.
Worked example: centre 10 mGy and edge 16 mGy give CTDIw = 10 ÷ 3 + 2 × 16 ÷ 3 = 3.3 + 10.7 = 14 mGy. At a pitch of 1.4, CTDIvol = 14 ÷ 1.4 = 10 mGy; at a pitch of 0.7 it would be 20 mGy.
It is proportional to the mAs per rotation. It rises steeply with the kV, faster than in proportion. It falls as the pitch rises, if the tube current is unchanged. It does not change with the length scanned. The measured profile includes overbeaming. Because it measures the scanner's output in a fixed object, it is good for comparing protocols and scanners and for reference levels. Lowering it in practice is lesson L58.
DLP = CTDIvol × scanned length, in mGy cm: the total radiation delivered. With CTDIvol at 10 mGy, a 45 cm abdomen and pelvis gives 450 mGy cm. The displayed DLP generally includes the extra length irradiated by overranging. An examination's total DLP is the sum of the DLPs of every series: each contrast phase and any repeat.
Effective dose E ≈ k × DLP, where k depends on the body region. The standard adult coefficients, from the European guidelines and tabled in AAPM Report 96, are about 0.0021 mSv per mGy cm for the head, with a DLP from the 16 cm phantom, and 0.014 for the chest and 0.015 for the abdomen and pelvis, with a DLP from the 32 cm phantom. Children have larger coefficients.
So 450 mGy cm × 0.015 is about 7 mSv. A head at CTDIvol 50 mGy over 16 cm gives 800 mGy cm and about 1.7 mSv, close to L42's 2 mSv. L42's typical figures, about 2 mSv for a head, 8 for a chest and 10 for an abdomen and pelvis, are older; newer figures are lower. The estimate is for a reference person of standard size: good for comparing examinations and for explaining risk in broad terms, not for one patient's risk.
The phantom is one size; patients are not. With the same technique, a small adult or a child absorbs more than the 32 cm phantom's figure, because less tissue surrounds the middle; a large adult absorbs less. The size-specific dose estimate, from AAPM Reports 204 and 220, corrects for this: it is CTDIvol × a factor found from the patient's size. Report 204 sized the patient by measured width; Report 220 by water-equivalent diameter, from the attenuation on the localiser. For a 32 cm figure the factor is about 1.8 for a patient 20 cm across, about 1 at 35 cm and about 0.85 at 40 cm.
For the same technique, CTDIvol against the 16 cm phantom is about twice that against the 32 cm one. Scanners have given children's body scans against either, so read the phantom off the report before comparing. And CTDIvol is a phantom average: it says nothing about the dose to a particular organ, such as the lens or the breast.
UK national diagnostic reference levels (UKHSA, list of December 2025, revised from time to time) give CTDIvol per sequence and DLP per examination, generally the third quartile of a national survey. Adults: head for acute stroke about 47 mGy and 790 mGy cm (16 cm phantom); chest for lung cancer about 8.5 and 290; CT pulmonary angiogram about 9.1 and 310; abdomen and pelvis about 10 and 530; kidneys, ureters and bladder about 6.3 and 290 (32 cm phantom). Reference levels in law are lesson L47; safety in the CT room, L61.
A typical abdomen and pelvis: the localiser, with a very small CTDIvol and DLP; then a helical series, phantom Body 32, CTDIvol 10 mGy, DLP 470 mGy cm. Total DLP a little over 470 mGy cm, and 470 × 0.015 is about 7 mSv. A second phase adds its DLP to the total; its CTDIvol is not added.
CTDIvol does not change with scan length; the DLP does. CTDIvol is the dose to a phantom: smaller patients absorb more, larger ones less. Add DLPs, not CTDIvols. Effective dose from the DLP is an estimate for a reference person, not one patient's risk.