Spins, precession and resonance

Magnetic resonance Free lesson Lesson 62 of 90

Proton spin, B0, Larmor equation, net magnetisation M0 vs field strength, B1 excitation and flip angle, Mz/Mxy, FID; magnet and RF-coil overview

0:00 / 0:00

Do you like iRad Shots?

Tap a star

INDEX

The shot

In one line

The magnetic resonance signal comes from hydrogen nuclei. A strong field gives them a small net magnetisation; a radio pulse at their precession frequency tips it over; as it turns, it induces a voltage in a coil.

Know this
  • The hydrogen nucleus is a single proton: it spins, so it acts as a tiny magnet. Hydrogen is used because it is plentiful in water and fat, with a strong magnetic moment.
  • With no field the protons point every way, and cancel. In the main field, B-nought, they line up with it or against it, a few per million more with it.
  • This excess gives a small net magnetisation, M-nought, along B-nought. It grows in proportion to the field: one reason a higher field gives more signal.
  • The protons precess round B-nought like spinning tops, at the Larmor frequency: the gyromagnetic ratio times the field: about 64 megahertz at 1.5 tesla, and 128 at 3 tesla.
  • A radiofrequency pulse, B-one, at right angles to B-nought and at exactly the Larmor frequency, brings the protons into step and tips the net magnetisation over. Off resonance, nothing happens.
  • The flip angle grows with the pulse's strength and length. 90 degrees lays the net magnetisation in the transverse plane; 180 degrees turns it upside down.
  • Longitudinal magnetisation cannot be measured. Transverse magnetisation, rotating at the Larmor frequency, induces a voltage in a coil: the free induction decay, which fades as the protons fall out of step.
  • The scanner: a superconducting magnet in liquid helium, always on; a body coil to transmit, receive coils close to the patient, gradient coils, and a room shielded from radio waves.
Numbers
  • Hydrogen's gyromagnetic ratio: about 42.6 megahertz per tesla.
  • The excess with the field: about 5 per million at 1.5 tesla, 10 at 3 tesla.
  • Liquid helium: about 4 kelvin.
  • The main field is kept even to within a few parts per million.
Exam traps
  • The pulse tips the net magnetisation. It does not push every proton into the high-energy state.
  • Only transverse magnetisation is measured.
  • Double the field, and the Larmor frequency doubles.
  • A pulse off the Larmor frequency does nothing.

Sketch

Hand-drawn sketch: Spins, precession and resonance

Check yourself

As in the exam: mark each item true or false. Any number can be true.

1. Concerning the hydrogen nucleus in MRI:

  1. aIt behaves as a small magnet because it has spin.
  2. bWith no external field, the protons in tissue line up to give a net magnetisation.
  3. cIt is used mainly because it is the heaviest nucleus in the body.
  4. dIt is used because it has no magnetic moment.
  5. eIt is a single proton.

2. Regarding the net magnetisation M0 in the main field B0:

  1. aAt equilibrium it has a large transverse part.
  2. bIt comes from about half of the protons lining up with the field.
  3. cIt points along B0.
  4. dIt is proportional to the field strength.
  5. eThe lower-energy state is the one lined up against B0.

3. Concerning precession and the Larmor frequency:

  1. aFor hydrogen at 1.5 tesla it is about 64 megahertz.
  2. bThe Larmor frequency is proportional to the field strength.
  3. cAt 3 tesla it is about 32 megahertz.
  4. dDoubling the field halves the Larmor frequency.
  5. eThe gyromagnetic ratio is the same for every nucleus.

4. Regarding the radiofrequency pulse:

  1. aIt tips the magnetisation at any frequency, provided it is strong enough.
  2. bA longer pulse gives a larger flip angle.
  3. cA 90-degree pulse pushes all the protons into the high-energy state.
  4. dIt brings the precessing protons into step.
  5. eIts magnetic field, B1, is at right angles to B0.

5. Concerning the signal and the scanner:

  1. aThe free induction decay oscillates at the Larmor frequency.
  2. bThe main magnet of a superconducting scanner is switched off between patients.
  3. cThe signal comes from the transverse magnetisation rotating past a receiver coil.
  4. dThe scanner room is shielded against radio waves.
  5. eLongitudinal magnetisation is measured directly.

More questions in the question bank
The full explanation about 5 min to read

Where the signal comes from

MRI rests on one chain of cause and effect. A strong magnetic field gives the hydrogen nuclei in the body a small net magnetisation. A radio pulse at exactly the right frequency tips that magnetisation away from the field. Once tipped, it turns round the field, and a turning magnet induces a voltage in a nearby coil. That voltage is the signal; the rest of this module is about how it is shaped, located and turned into an image.

The proton as a magnet

The nucleus of ordinary hydrogen is a single proton. It has a property called spin, and because it is a spinning charge it behaves like a tiny bar magnet, with a magnetic moment. Hydrogen is used in clinical MRI for two reasons: it is by far the most plentiful nucleus in the body, in water and in fat, and its magnetic moment is strong, so it gives a large signal.

Lining up in the main field

With no field, the protons point every way and their magnetism cancels: the tissue as a whole is not magnetised. In the scanner's strong static field, called B0, each proton takes up one of two states: lined up with the field, the lower-energy state, or against it, the higher-energy state. Heat keeps the two almost equal, so the excess with the field is tiny: about 5 per million at 1.5 tesla and about 10 per million at 3 tesla, at body temperature. But there are so many protons that the excess adds up to a net magnetisation, M0, along B0.

M0 is proportional to the field strength: double the field and the excess doubles. That is one of the main reasons a higher field gives more signal; signal-to-noise is taken further in L67.

Precession and the Larmor equation

Like a spinning top leaning in gravity, a proton's axis wobbles round the direction of the field. This is precession, and its frequency is the Larmor frequency, set by the Larmor equation: frequency equals the gyromagnetic ratio times the field strength. For hydrogen the gyromagnetic ratio is about 42.58 megahertz per tesla, so the frequency is about 63.9 megahertz at 1.5 tesla and about 127.7 megahertz at 3 tesla. These are radio frequencies: non-ionising (L01).

At rest, each proton precesses with its own random phase, so their sideways components cancel and M0 points straight along B0, with nothing left in the transverse plane.

Resonance and the flip angle

To get a signal, M0 must be tipped. A radiofrequency coil sends a short pulse whose magnetic field, called B1, is at right angles to B0 and rotates at exactly the Larmor frequency. Keeping in step with the precessing protons, even a weak B1 of a few microtesla steadily pushes the net magnetisation away from B0. This is resonance, like pushing a swing at its own rhythm. The pulse also brings the protons into step, so that their sideways components add instead of cancelling. A pulse at the wrong frequency drifts out of step and does nothing useful.

In the quantum picture, the energy of the radio photons matches the gap between the two states. But the classical picture of a tipping net magnetisation is the one that describes what is measured, and it should be used: the pulse rotates the net magnetisation; it does not simply flip each proton, nor push them all into the high-energy state.

The angle the net magnetisation is tipped through is the flip angle. It grows with the strength of B1 and with the length of the pulse. A 90-degree pulse lays the whole of the net magnetisation in the transverse plane; a 180-degree pulse turns it upside down, pointing against B0.

Longitudinal and transverse magnetisation

The net magnetisation is described in two parts. The longitudinal part, Mz, lies along B0. It cannot be measured directly, because it is tiny beside B0 and does not change. The transverse part, Mxy, lies at right angles to B0 and turns round it at the Larmor frequency. That rotating magnetisation can be measured.

The free induction decay

A receiver coil placed near the patient has the rotating transverse magnetisation sweeping past it, so, by Faraday's law, a voltage is induced: an oscillation at the Larmor frequency. After a single pulse this signal is called the free induction decay. It dies away within tens of milliseconds in tissue, because the protons fall out of step with one another, while the longitudinal magnetisation slowly grows back. The time constants of those two processes, T1, T2 and T2-star, belong to L63.

The scanner

The main magnet is a coil of superconducting wire cooled by liquid helium to about 4 kelvin, minus 269 degrees Celsius. With no resistance, the current flows without a power supply, so the field is always on (safety and the quench: L77 and L78). Clinical scanners work at 1.5 or 3 tesla, tens of thousands of times the Earth's field. The field must be very even, to within a few parts per million across the imaging volume, and shimming, with metal pieces and extra coils, makes it so.

Radiofrequency coils transmit and receive. A large body coil built into the bore usually transmits. Receive coils, often arrays of small elements, are placed close to the part being examined, because a coil near the tissue picks up more signal and less noise (L67). The gradient coils add small, controlled changes in the field to locate the signal (L65). The scanner sits in a radiofrequency-shielded room, a copper Faraday cage, that keeps outside radio signals from swamping the faint signal from the patient.

Exam traps

One: the pulse tips the net magnetisation; individual protons are not simply flipped, nor all pushed into the high-energy state. Two: only the transverse magnetisation is measured. Three: the Larmor frequency is proportional to the field, about 64 megahertz at 1.5 tesla and 128 at 3 tesla. Four: a pulse off resonance does nothing.